ANGIO-SEALĀ® (VIP Vascular Closure Device)
PRODUCT OVERVIEW

BIOABSORBABLE COMPONENTS
The device creates a mechanical seal by sandwiching the arteriotomy between a bioabsorbable anchor and collagen sponge, which dissolve in 60 to 90 days.
Three bioabsorbable components actively seal the arteriotomy:
- Anchor: Bioabsorbable co-polymer anchor placed against the inside of the vessel wall
- Collagen: Placed on top of the arteriotomy in the tissue tract
- Suture: Cinches the anchor and collagen together to form a secure seal.
LOCATE THE ARTERY
- Exchange the procedure sheath with the Angio-Seal locator system.
- Blood flow through the locator visually confirms proper sheath position in the artery.
SET THE ANCHOR
- Insert the Angio-Seal VIP device into the sheath until you hear a āclick.ā
- Gently pull back on the locking cap until you hear another āclick.ā
- The anchor is now locked in place and device is ready to be deployed.
SEAL THE PUNCTURE
- Gently pull back on the Angio-Seal VIP device until the suture has stopped spooling.
- Maintain upward tension on the device and gently advance the compaction tube until resistance is felt.
- Cut the suture and remove the device.
ANGIO-SEAL VIP PRODUCT CODES
| Part Number | French Size | Guidewire Diameter (in) |
| 610130 | 6Fr | .035 Wire |
| 610131 | 8Fr | .038 Wire |
REFERENCES
RX ONLY. Refer to the product labels and package insert for complete warnings, precautions, potential complications, and instructions for use.Ā
- Nash JE, Evans DG. (1999). The AngioāSeal hemostatic puncture closure device. Concept and experimental results. Herz, 24(8), 597ā606.
- Per Instructions For Use ASIN0004 revision 2018-09-01.
Indications:
TheĀ Angio-Seal Vascular Closure Device product family, including the VIP and Evolution platforms, is indicated for use in closing and reducing time to hemostasis of the femoral arterial puncture site in patients who have undergone diagnostic angiography procedures or interventional procedures using an 8 French or smaller procedural sheath for the 8 F Angio-Seal device and a 6 French or smaller procedural sheath for the 6 F Angio-Seal device.Ā The Angio-Seal VIP and Evolution platform devices are also indicated for use to allow patients who have undergone diagnostic angiography to safely ambulate as soon as possible after sheath removal and device placement, as well as to allow patients who have undergone an interventional procedure to safely ambulate after sheath removal and device placement.
Important Safety Information:
Possible adverse events for vascular closure devices include, but are not limited to: bleeding or hematoma, AV fistula or pseudoaneurysm, infection, allergic reaction, foreign body reaction, inflammation or edema. This device should only be used by a licensed physician (or other health care professional authorized by or under the direction of such physician) possessing adequate instruction in the use of the device, e.g., participation in an Angio-Seal physician instruction program or equivalent.
ANGIO-SEALĀ® (VIP Vascular Closure Device)
Related products
-
RADIFOCUS REDIAL SHEATH
Providing you the best range of Terumo Radifocus Radial Introducer Sheath and Cordis Femoral Introducer Sheath with effective & timely delivery.
The radial artery access sheath isĀ used to gain arterial access and facilitate the insertion of catheters or other equipment for diagnostic and vascular interventions.
- The Cordis AVANTIĀ®+ Introducer is the pioneer of catheter sheath introducer technology. Featuring a patented SLIX Valve, the AVANTIĀ®+ introducer provides smooth transitions, monitoring capabilities and exceptional performance for your procedural success, as well as
- A hexacuspid design that provides a balance between catheter maneuverability and hemostasis.
- A rotating suture collar that facilitates procedural flexibility. It stays in place and allows patient movement.
- A kink-resistant cannula design integrating a soft, flexible inner layer with a stiffer outer layer designed to increase bendability and support, and decrease kinking.
- Atraumatic tip transitions for both the sheath and the vessel dilator are uniquely tapered and manicured. This results in smooth insertions and helps to minimize damage upon entry.
As part of the circulatory system,Ā the radial artery supplies blood from the heart to the forearm.
Radial sheaths had the smallest OD (1 Fr smaller) when compared to similar sized conventional sheaths. For instance, the OD of 4 Fr radial sheath (1.610 ± 0.006 mm) is essentially the same as the OD of the conventional 3 Fr (1.644 ± 0.016 mm) and 3.3 Fr (1.635 ± 0.005 mm) sheaths.
-
NC Sprinter PTCA BALLOON
NC Sprinter RX Noncompliant
NC Stormer over-the-wire noncompliant balloon dilatationĀ catheter offers controlled compliance, high pressures, and comprehensive lengths.
Balloon Dilatation Catheter
OVERVIEW
The NC Sprinter⢠RX noncompliant coronary balloon combines:
- Sprinter crossing technology
- High-pressure capability and controlled expansion1
ORDERING INFORMATION
BALLOON LENGTHS 6, 9, 12 MM
Balloon Diameter (mm)
Balloon Length (mm)Ā Ā Ā Ā Ā 6 9 12 2.00 NCSP2006X NCSP2009X NCSP2012X 2.25 NCSP22506X ā NCSP22512X 2.50 NCSP2506X NCSP2509X NCSP2512X 2.75 NCSP27506X NCSP27509X NCSP27512X 3.00 NCSP3006X NCSP3009X NCSP3012X 3.25 NCSP32506X NCSP32509X NCSP32512X 3.50 NCSP3506X NCSP3509X NCSP3512X 3.75 NCSP37506X NCSP37509X NCSP37512X 4.00 NCSP4006X NCSP4009X NCSP4012X 4.50 ā ā ā 5.00 ā ā ā BALLOON LENGTHS 15, 21, 27 MM
Balloon Diameter (mm)
Balloon Length (mm)Ā Ā Ā Ā Ā 15 21 27 2.00 NCSP2015X NCSP2021X ā 2.25 ā NCSP22521X ā 2.50 NCSP2515X NCSP2521X NCSP2527X 2.75 NCSP27515X NCSP27521X ā 3.00 NCSP3015X NCSP3021X NCSP3027X 3.25 NCSP32515X NCSP32521X ā 3.50 NCSP3515X NCSP3521X NCSP3527X 3.75 NCSP37515X NCSP37521X ā 4.00 NCSP4015X NCSP4021X NCSP4027X 4.50 NCSP4515X NCSP4521X ā 5.00 NCSP5015X ā ā -
EXPORT ADVANCE ASPIRATION CATHETER
EXPORT ADVANCE ASPIRATION CATHETER
Export Advance
Aspiration Catheter
ADVANCING PERFORMANCE
The Export Advance⢠aspiration catheter delivers consistent, high-performing aspiration power when it matters most ā restoring flow and protecting patients.
PRODUCT DETAILSā
Superior Deliverability1
Full-Wall TechnologyĀ provides variable levels of stiffness without joints, for optimalĀ kink resistance and catheter performance.
Dependable Delivery to the Target Site:
- Preloaded stylet enhances shaft stiffnessĀ during catheter delivery, boosting kink resistance, trackability, and pushability2
- Buddy wire compatibilityĀ allows for extra support, when needed.ā
High-performing Aspiration Power:
- Large extraction lumen (0.044″ proximal; 0.043″ distal) increases aspiration power.ā”
- Optimised hub geometry improves flow.ā”
- Soft, short, forward-facing tip design permits excellent particle capture.ā
ORDERING INFORMATION
ADVANCE 6 F min. guide I.D. 0.070″ 0.014 140 - ā¢Third-party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company.
- Based on bench test data, 2013. Bench test data may not be indicative of clinical performance.
- Based on bench test data vs. Export⢠AP catheter.
- Bench testing vs. Terumo Eliminateā¢* and Eliminateā¢* 3 aspiration catheters, Vascular Solutions ProntoĀ®*Ā LP, ProntoĀ®*Ā V3 and ProntoĀ®*Ā V4 aspiration catheters, MeritĀ®*Ā ASAP aspiration catheter, Kaneka Thrombuster III GR* aspiration catheter, Spectranetics QuickCatā¢* aspiration catheter, IHT Polska HunterĀ®* aspiration catheter, Medrad Possisā¢* Fetchā¢* 2 aspiration catheter, and Hexacath Recover* aspiration catheter. 10093906DOC, 2013. Bench test data may not be indicative of clinical performance.
- Based on animal studies compared with Export⢠AP aspiration catheter without stylet.
-
Sprinter OTW Balloon
Sprinter OTW Semicompliant
Balloon Dilatation Catheter
Sprinter⢠over-the-wire semicompliant balloon offers low-tip and crossing profiles with a comprehensive size matrix
OVERVIEW
The Sprinter⢠over-the-wire (OTW) semicompliant balloon dilatation catheter is available in a broad size matrix.
FEATURES
MiniWrap folding is used on 1.50-4.00 balloons. The 1.50 mm balloon has two folds.
- Ā Selective Dura-Trac coating
- Ā 2.5 mm tip
- Ā FasTrac tip
Balloon’s
- Distal shaft
- Gold-swaged marker bands
- Selective Dura-Trac hydrophilic coating
- Fulcrum balloon material
- Proximal shaft (oval)
- FasTrac tip








Reviews